06 07 Wrapped intruments Program P1 Program P2 Program P3 Program P4 Program PRION E7 five cycles for unequalled safety With the simple press of your finger you can select in an ergonomic and intuitive manner the sterilization cycle which is more

Transcription:

the autoclaves with experience,Safe and tested practices. appreciated every day, The Euronda E7 and E5 autoclaves The E7 and E5 autoclaves are easy. are safe thanks to the PED and intuitive to use thanks to the. 97 23 EEC certified stainless steel simple and clear commands and. sterilization chamber and equipped the extremely ergonomic. with an absolutely safe closing open close system The multi. system The final sand polishing language LCD display with which. treatment guarantees better the E7 is equipped assists you step. resistance and makes the surfaces by step always supplying you with. more easy to clean the correct information for an,effective use. the complete autoclave, Behind the shiny 18 or 24 litre PED certified stainless steel chamber lies first. class equipment top of which are the double head vacuum pump and the. sophisticated microprocessor which constantly monitors the key parameters. for a correct sterilization such as temperature pressure and vacuum The. two separated water tanks avoid the contamination between clean water to. be used for sterilization and the used water The external thermal paper. printer makes its use even more pleasant and practical. E7 five cycles,for unequalled safety, Unwrapped intruments Unwrapped intruments Wrapped intruments Wrapped intruments Prion cycle.
With drying With drying With drying With drying, Program P1 Program P2 Program P3 Program P4 Program PRION. E7 18 l E7 24 l E7 18 l E7 24 l E7 18 l E7 24 l E7 18 l E7 24 l E7 18 l E7 24 l. Material Solid unwrapped Material Solid unwrapped Material Solid wrapped Material Solid wrapped Material Solid wrapped. Temperature 121 C Temperature 134 C Temperature 121 C Temperature 134 C Temperature 134 C. Duration of Duration of Duration of Duration of Duration of. sterilization 20 sterilization 4 sterilization 30 sterilization 12 sterilization 18. phase phase phase phase phase, Total time 63 Total time 49 54 Total time 69 68 Total time 53 Total time 59. With the simple press of your finger you can select in an ergonomic and intuitive manner the sterilization cycle which is more. adapt for your need The quality of the vapour generated will be able to guarantee you an effective and safe sterilization. independently from the chosen cycle and the instructions given by the LCD display will make its use practical and pleasant. the simple autoclave, The E5 autoclave maintains the quality E5 is the class N fully automatic. characteristics of its bigger sister E7 even microprocessor controlled electronic. with simpler equipment and performances autoclave characterised by a. It is available only in the 18 litres version thermodynamic working method with. and is equipped with the essential control of the pressure and. commands which include the manometer temperature E5 is particularly. with which to control the correct sterilization user friendly thanks to the simple. pressure The Euronda E5 is also equipped and intuitive control panel that can. with the PED certified chamber two be used to select between two. separated water tanks and door close sterilization cycles 121 C and 134 C. safety system and the LEDs that signal the various. phases of the cycle being carried out,E5 choose the cycle. leave the rest to her,Program 121 Program 134,E5 18 l E5 18 l.
Material Solid unwrapped Material Solid unwrapped,Temperature 121 C Temperature 134 C. Duration of Duration of,sterilization 20 sterilization 10. phase phase,Total time 57 Total time 52, Choose the sterilization cycle which better suits your needs On the command panel you may follow the state. through acoustic and visual signals with the certainty that in any case the result will be the correct sterilization. of the instruments inside the autoclave for the safety of yourself your staff and your patients. E7 18 Overall space needed E7 18 Usable space,Space required for the movement of the door. E7 24 Overall space needed E7 24 Usable space,Space required for the movement of the door.
TECHNICAL SPECIFICATIONS OF THE AUTOCLAVE E7 Safety devices. Closing procedure by safety microswitch,SPECIFICATIONS E7 18 l E7 24 l. Door locking system by safety,microswitch, Electric supply 230V Resistance with double safety thermostat. Safety valve, Frequency 50 Hz Complete self test of the sterilization. Power consumption 1700 W 2000 W working process,Absorbed current 7 4 A 8 7 A. Sterilization programs 5 Standards,EN 61010 1,Test cycles Vacuum test EN 61010 2 040.
Indoor device EN 61326,Altitude up to 2000 from sea level EN ISO 14971. Operational range,Temperature 5 40 C EN 13445,Relative humidity Max 85. Max variation of the net voltage 10,Class of installation class of overvoltage II. Degree of pollution 2,Certifications, Max relative pressure 250 kPa 2 5 bar Autoclave E7. Conforms to the Directive, Size of the sterilization chamber 250 mm Depth 370 mm 250 mm Depth 470 mm.
93 42 EEC and following, Volume of the chamber 18 l 24 l modification 2007 47 CE. Medical Device Class II A, Usable space 180 x 160 x 280 mm LxHxD 180 x 160 x 380 mm LxHxD Chamber conform to the Directive. Usable volume 8l 11 l 97 23 CEE, External size of the autoclave 450 x 385 x 500 mm LxHxD 450 x 385 x 600 mm LxHxD ISO 9001. Empty 39 7 Kg Empty 44 2 Kg, Weight of the autoclave Max weight with full tank Max weight with full tank and. and max load 55 Kg,max load 48 Kg,Clean water tank capacity 3l 3l.
Controlling system Electronic via microprocessor,Max noise reveled 58dB A. Bactereological filter Yes,E5 Overall space needed E5 Usable space. Space required for the movement of the door, TECHNICAL SPECIFICATIONS OF THE AUTOCLAVE E5 EN13060 Safety devices. Closing procedure by safety microswitch,SPECIFICATIONS E5 18 l. Door locking system by safety,microswitch, Power supply 230V Resistance with safety thermostat.
Safety valve, Frequency 50 Hz Complete self test of the sterilization. Power consumption 1500 W,Absorbed current 6 5 A,Sterilization programs 2 EN 61010 1. EN 61010 2 040,Indoor device EN 61326,Altitude up to 2 000 from sea level EN ISO 14971. Temperature 5 40 C, Operational range Relative humidity Max 85 EN 13445. Max variation of the net voltage 10,Installation class overvoltage class II.
Degree of pollution 2,Certifications, Max relative Pressure 250 kPa 2 5 bar Autoclave E5. Conforms to Directive, Size of the sterilization chamber 250 mm Depth 340 mm. 93 42 EEC and following,Volume of the chamber 18 l. modification 2007 47 CE,Medical Device Class II A, Usable space 180 x 160 x 280 mm LxHxD Chamber conform to the Directive. Usable volume 8l Company, External size of the autoclave 450 x 385 x 500 mm LxHxD ISO 13485.
Weight of the autoclave Empty 30 35 Kg, Max weight with full tank and completely loaded 35 43 Kg. Clean water tank capacity 3l,Controlling system Electronic via microprocessor.